Summary
GD-3536 Manual Flash and Fire Point Tester is designed and made as per GB/T3536-2008 Petroleum Products - Determination of Flash and Fire Points - Cleveland open cup method and ASTM D 92 Standard Test Method for Flash and Fire Points by Cleveland Open Cup Tester. It is applicable to all petroleum products with flash points above 79°C and below 400°C except fuel oils.
Main technical specifications and parameters
1 Power supply AC(220±10%)V,50Hz
2 Heating device Quartz tube furnace heating, no naked fire, explosion prevented.The power is adjustable from 0W to 600W
3 Test flame applicator: It applies the test flame automatically
4 Thermometer (-6~400)°C. It is the same as thermometer ASTM 11C
5 Igniting device
(1) Ignition source: coal gas(or civil gas,the same below)
(2) Flame diameter is 3.2mm~4.8mm
6 Ambient temperature (-10~50)°C
7 Relative humidity =85%
8 Overall power consumption = 650W
9 Overall dimension 350mm290mm350mm (thermometer is not included)