The UV aging test chamber is suitable for all kinds of geo-synthetics, such as light resistance, weather resistance and light ageing
The exposure test of the sample was carried out in a fluorescent UV lamp vessel in a controlled environment. It can simulate and strengthen the environmental factors in the natural climate that are affected by light, heat, humidity, and rainfall as the main aging. It can simulate the accelerated test under long term exposure to obtain the results of the weather resistance of the material.
· JTG E50 T1164 The methods of geo-synthetics for highway engineering
· GB/T 14522-93 Accelerated weathering test method for plastics, coatings and rubber materials used for machinery industrial products.
· GB-T16422.3-1997 PlasticsMethods of exposure to laboratory light sourcesPart 3: Fluorescent UV lamps
· ISO4892-3:1994 Plastics - Methods of exposure to laboratory light sources - Part 3: Fluorescent UV lamps
Standard Structures
· Main features: light, condensation, automatic cycle control, real-time monitoring and display
· Light temperature: RT + 10 ℃ ~ 70 ℃
· Condensation temperature: 50 ℃
· Condensation method: steam humidification condensation
· Temperature uniformity: ± 3 ℃
· Irradiation intensity: 0.55W / m2
· Light / condensation cycle: 0 ~ 9999h,59min. It can be set (such as: 4H / 4H or 8H / 8H)
· Lamp center distance: 70mm
· Sample and lamp plane distance: 55mm
· UV lamp: UV-A340 or UV-B313 (optional)
· UV wavelength: (UVA) 315 ~ 400nm; (UVB) 280 ~ 315nm
· Lamp life: 1600 ~ 1800H
· Standard test piece: 75 × 150mm or customized according to customer requirements (it can be made for 21 samples)
· Control system: touch screen controller, real-time monitoring friendly interface, easy to use and easy to operate.
· Power supply: AC220V 50HZ 4.7KW
Technical Parameters
Features
Stable reliable and consistent
· The structure of the test chamber is made of corrosion-resistant metal material, including 8 fluorescent UV lamp, water tray, test sample rack and touch screen controller.
· Lamp power is 40W and lamp length is 1200 mm. The working area of the test chamber is in the range of 900 × 210 mm.
· The lamp is installed in a row of four rows, installed in two rows, and the lamp is arranged in parallel, and the center distance of the lamp is 70 mm.
· The test sample is fixedly mounted at a distance of 50 mm from the nearest parallel surface of the lamp surface. The test specimen and its support form the inner wall of the box, and their backs are exposed to room temperature cooling air due to the temperature difference between the test sample and the air in the box. So that the test sample surface condensation phase to produce a stable condensation conditions, the test chamber shall be of natural air convection from the bottom of the box